Technical Parameter
| Measurement range: | 0 to 10,000 ppm O₂ | Operating pressure range | 0.5-2bar absolute |
| Electrical connector: | 2xslip-rings on PCB | Pressure dependency | Partial Pressure |
| Initial output signal: | 460μA±120μA ambient air | Operating temperature | 0 To 50℃ |
| Response times(in op- eration): | <180 s from 10,000 ppm | Flow rate | 0-2.5 Lpm |
| down to 1,000 ppm | Temperature compensation | None | |
| <10 min from 1,000 ppm | Weight | Approximately 35g | |
| down to250 ppm | Storage duration: | <3 months | |
| Linearity error: | ±2 % of signal | Sensor lifetime: | 1 year, depending on |
| Drift (% signal month): | <1% | humidity and O₂ | |
| Humidity non-con- densing: | 0-99% RH | Material in contact with Media: | PV DF, PT FE, stainless steel |
| Linearity: | ± 1% FS | Expected life: | 24 Months |
Note: Use sensor immediately after removing from the packaging. After removing, do not leave the sensor in air for an extended period of time. Failure to do so may have a negative impact on its performance and life.
Oxy-550-XL Replaces: Teledyne B-2CXL, XLT-12-333, OX-2