CI-PC126 Frequency conversion ion flow Oxygen Analyzer Product Features
Data storage function: real-time record test data and canquickly query the stored data;
When the oxygen concentration deviates from the alarmvalue, the alarm value can be set arbitrarily in the full range;
Gas path pressure detection function: real-time detection ofanalyzer gas path pressure, easy to judge whether the gaspath isblocked;
RS485 two-way communication function: can communicate with the computer or other digital communications equipment directly.
Technical Index
| Sensor | Frequency conversion ion flow sensor | |
| Measuring range | 0~25% O₂ | |
| Linear error | 0.01~99.9ppm | ±10ppm |
| 100~999.9ppm | ±3%FS | |
| 1000~9999ppm | ±3%FS | |
| 1.00%~25.00% | ±3% of measured value | |
| Repeatability | 0.01~99.9ppm | ±5ppm |
| 100~999.9ppm | ±1.5%FS | |
| 1000~9999ppm | ±1.5%FS | |
| 1.00%~25.00% | ±1.5% of measured value | |
| Stability | 0.01~99.9ppm | ±5ppm/7d |
| 100~999.9ppm | ±1.5%FS/7d | |
| 1000~9999ppm | ±1.5%FS/7d | |
| 1.00%~25.00% | ±1.5%/7d of measured value | |
| Response time | T₉₀<60s(after full preheating) | |
| Ambient TEMP. | -10~+50℃ Ambient humidity:<80%RH | |
| Sample gas flow | 400~600ml/min | |
| Analog signal | 4-20mA/0-20mA/0-1V/0-5V/0-10V | |