The sales of the CHANG AI also contributes to the quality of the service.
CI-PC86 High content Oxygen Analyzer Product Features
Adopting capacitive touch buttons, the buttons are sensitive, reliable, and easy to operate;
Built in one-way pump to prevent the analyzer from turning on in high oxygen environments and extend the service life of the sensor;
Menu lock switch: to prevent the impact of misoperat-
ion on thenormal operation of the analyzer;
When the oxygen concentration deviates from the alarm value, the alarm value can be set in the full range of arbitrary.
Technical Index
Sensor | 3D ion flow sensor |
Measuring range | 10~97.00% / 99.99%, 97~99.99% O₂ |
Linear error | ±1%FS |
Repeatability | ±0.5%FS |
Stability | ±0.5%FS/7d |
Response time | T₉₀<15s |
Power supply | AC 100~240V, 50/60Hz |
Consumption | <20VA |
Communication | RS485(standard)/RS232(optional) |
Ambient TEMP | -10~+55℃ |
Ambient humidity | <85%RH |
Measurement method | Gas inlet type |
Sample flow | 400~600ml/min |
Flow control | Built-in needle valve |
Dimensions | L×W×H=144mm×144mm×290mm |
Installation | Disk mounted |
The increasing demand of customers to CHANG AI also helps to be more competitive in this industry.