The quality of the product can stand the test of time.
CI-PC983 Oxygen transmitter Product Features
Using electrochemical sensor as measuring unit, high precision, long calibration period, stable signal output;
The sensor does not need external power supply and heating, the whole machine power is extremely low;
The sensor is not affected by CO2, CO, H2S, NOX, H2;
Communication function: use RS485(standard)/RS232(optional) two-way communication, can communicate directly with computer or other digital communication equipment;
Isolated analog signal output, support software switching 0-20mA/4-20mA analog output mode.
Technical Index
| Sensor | Electrochemical oxygen sensor |
| Measuring range | 0~25.00% O2 |
| Linear error | ±2%FS |
| Repeatability | ±1%FS |
| Stability | ±1%FS/7d |
| Response time | T90<20s |
| Power supply | DC 24V, 0.5A |
| consumption | <10VA |
| Communication | RS485 (standard)/RS232 (optional) |
| Analog signal | Number 1 group |
| Output mode 4-20mA standard/0-20mA optional load <500Ω | |
| Ambient TEMP. | -10~+45℃ Ambient humidity<90%RH |
| Sample gas flow | 0.3~1L/min |
| Sample gas pressure | 0.3-0.6 mpa, 0.5 MPA recommended (outlet shall be at atmospheric pressure) |
| Air lntakes | 1/4 PT thread, standard 6.5 quick screw |
| Installation | “几” type bracket installation or“L” type bracket installation |
CHANG AI is manufactured under rigorous materials inspection procedure and standardized production conditions.