Product Advantages
The CI-PC631 In-situ laser trace humidity analyzer features high measurement accuracy, fast response time, high sensitivity, outstanding stability and reliability. It is equipped with analog signal output, data communication and other functional configurations, and supports the display of humidity volume ratio and dew point value (temperature). It is mainly applied to monitor the moisture content in glove boxes during the production of lithium batteries.
Product Features
PTFE Filter: Its inherent hydrophobicity ensures measurement accuracy while filtering out other impurities, thereby guaranteeing the transmittance of optical signals.
Prevents adhesion of viscous substances and keeps the mirror surface clean.
Utilizes KF40 standard flange for quick and convenient docking and installation.
Utilizes a multi-reflection path. Compared with the single-pass path, it provides a longer effective measuring optical path, which greatly improves measurement precision and accuracy.
The unique and compact gas absorption cell realizes ultra-small volume and exquisite appearance.
The internal surface of the absorption cell features a specially treated coating to ensure the lens with excellent corrosion resistance and wear resistance.
Built-in temperature sensor for real-time temperature compensation.
Specifications
| Principle | Non-Dispersive Infrared (NDIR) |
| Measurement range | 0-1000ppm H2O |
| Linear error | ±1%FS |
| Stability | ±1%FS /7d |
| Zero drift | ±0.1% FS /7d |
| Measurement response time | 3s |
| Ambient TEMP. | -10~60℃ |
| Ambient humidity | ≤90%RH |
| Analog output | 4-20mA/0-5V/0-10V |
| Communication output | RS485 |
| Power supply | 24VDC |
| power supply wattage | ≤10W |
| Gas inlet and outlet ports of the test chamber | Φ6 Plastic ferrule fitting(optional) |
| Wiring | Aviation connectors |
| Weight | About 1.85kg (including the sealed cover of the test chamber) |