This product has passed related qualifications and international certifications.
CI-PC25-1 Explosion-Proof Infrared Gas Analyzer Product Features
Using NDIR (that is, non-dispersive infrared measurement), the sensor is not in contact with the measured gas, and the poisoning of the sensor will not occur;
High stability infrared light source and filter with strong anti-interference ability make the analyzer have good long-term stability;
With microprocessor as the core, it has the characteristics of intelligence, good stability, high precision and long calibration cycle.
Technical Index
| Detection principle | NDIR non-dispersive infrared | |
| Component | Min range | Max range |
| CO₂ | 0~2000ppm | 0~100% |
| CO | 0~5000ppm | 0~100% |
| CH₄ | 0~3% | 0~100% |
| The measurement background gas must be ammonia, oxygen, hydrogen, etc. and the gas to be measured has no weight | ||
| Linear error | ±2% FS | |
| Repeatability | ±1% FS | |
| Zero drift | ±1% FS/7d | |
| Range drift | ±1% FS/7d | |
| Response time | T₉₀<15s(500ml/min) | |
| Power supply | DC 24V, 1A | |
| Consumption | <15VA | |
| Communication | RS485 (standard)/RS232 (optional) | |
| Ambient TEMP. | -10~+45℃ Ambient humidity:<90%RH | |
| Sample gas flow | 500~1000ml/min (best 500ml/min) | |
| EX-Mark | Ex d IIB T6 Gb | |
The product has passed all relative quality certificates.