CI-XT60EO-O₂X chemical explosion-proof online oxygen analysis system is a pretreatment system independently developed by our company. The system adopts anti-corrosion and explosion-proof treatment methods, which meet the requirements of explosion-proof production lines. The pretreatment system of this set of analysis system is composed of vacuum generator, automatic liquid drainage device, filter, calibration valve, analyzer instrument, cabinet and other components.
CI-XT60EO-O₂X Online Gas Analysis System Product Features
It has the characteristics of high measurement accuracy, fast response, high stability, strong reliability, and long calibration cycle;
Explosion-proof laser Oxygen Analyzer, one of the most advanced laser oxygen analyzers in the world;
The principle of laser optical detector is used to detect the oxygen content in the atmosphere;
Unique air circuit design, good air tightness, impermeability and water absorption resistance;
It has the function of interfering gas compensation to eliminate the influence of interfering gas on the measurement accuracy;
Analog output signal: 1 channel standard output 4-20mA.
Technical Index
Performance Parameters | |
Sensor | Laser optical detection principle |
Measuring range | 0~5% O₂ |
Linear error | ±0.2% O₂ |
Stability | ≤±1% reading/year |
Zero drift | ±0.1% O₂ /year |
Response time | Lag time:T₁₀≤5s |
Rise time:T₁₀~T₉₀≤2s | |
Work pressure | 0.8~1.4 bar |
Analog signal | 4~20mA |
Storage TEMP. | +10~+50℃ |
Analyzer life | More than 5 years (normal use) |
Working Conditions | |
Power supply | AC 220V 50Hz |
Ambient TEMP. | +15~+45℃ |
Relative humidity | ≤85%RH |
Sample gas TEMP. | Normal temperature (≤35℃) |
Sample gas pressure | 30 KPa |
Sample gas flow | ≤3L/min |
Dust concentration | Zero |
Pressure range | 0~0.2MPa |
Nitrogen purging | Requirements: no oil, no water, no dust |
Pressure: ≥0.5Mpa | |
Compressed air | 0.3~0.5Mpa |
Air flow | ≤0.5m/s |
Ambience | The environment shall not contain explosive, corrosive and other gases, and have good ventilation conditions |
Safety performance | Insulation resistance not less than 5MΩ |
Under the test voltage of 1.25KV, breakdown and arcing will not appear in 1min | |
Configuration And Output | |
Reaction speed | Lag time:T₁₀≤30s |
Rise time:T₁₀~T₉₀≤2s | |
Drift | Zero drift: ≤±1% FS/7d |
Range drift:≤±1% FS/7d | |
Load | ≤750Ω |
Analog signal | 4-20mA |
Sampling method | Ejection extraction |
System Cabinet Parameters | |
System cabinet | Outdoor stainless steel rainproof cabinet |
Specification | L×W×H=500×400×1300(mm) |
Ex-mark | Ex deII CT6 Gb |
Location of installation | Outdoor |