The Gas Analysis System (hereinafter referred to as the system) is a process analysis instrument (such as CO2, O2 analyzer) and sampling pretreatment device and its subsidiary application guarantee part (standard gas). Through on-the-spot investigation, detailed understanding of the on-site process and working conditions, and the correct matching and reasonable combination of the system design, the analytical instrument can well adapt to the special and complex process conditions of the sample gas to be tested.
CI-XT800 CO₂ recycling online monitor system Product Features
| Reaction Speed | |
| Analyzer lag time (T10) and rise time (T10~T90.) | |
| Lag time | T₁₀≤5s |
| Rise time | T₁₀~T90≤2s |
| Lag time and rise time of the system | |
| Lag time | T₁₀≤30s |
| Rise time | T10~T₁₀≤2s |
Technical Index
Technical Index
| Main Technical Index | |
| Zero drift | ≤±1% FS/7d |
| Range drift | ≤±1% FS/7d |
| Repeatability | ≤0.5% |
| Load | ≤500Ω |
| Analog signal | DC 4~20 mA |
| Applicable Working Conditions | |
| Gas sample TEMP. | ≤60℃ |
| Gas sample flow | ≤300L/h |
| Gas sample dust concentration | ≤ 2000g/m3 |
| Corrosiveness | Gas sample should not contain strong corrosive components |
| Working Conditions | |
| Ambient TEMP. | +5~+45℃ |
| Relative humidity | ≤85% |
| Power supply | AC 220V±10% 50/60Hz |
| Consumption | ≤0.5KVA |
| Ambient air pressure | Altitude below 2500m |
| No direct sunlight radiation | |
| No strong electromagnetic field interference | |
| No mechanical vibration | |
| Air velocity | ≤0.5m/s |
| Working position | Horizontal |
| Quality System Certification: ISO9001: 2015 Standard | |