loading

Solid Electrolyte Zirconia Ion Flow Detection Technology

Introduction to Zirconia Analyzer Detection Technology and Ion Flow Sensors

With the development and maturation of zirconia sensor technology, the applications of zirconia sensors have expanded from automotive exhaust emission testing to industries and fields such as heating boiler control, industrial process control, combustion systems, oxygen/nitrogen generation systems, agricultural composting, and flue gas emissions. The analytical objects of zirconia sensors have also extended from the analysis of simple oxygen concentration to that of nitrogen oxides concentration, water vapor concentration, sulfur dioxide concentration and more. Nowadays, zirconia sensors have become one of the most important and widely used sensors in the field of gas analysis.

>> Zirconia Analyzer Detection Technology

The material used in zirconia sensors is a zirconia solid electrolyte. It is fabricated by doping pure zirconia with a certain proportion of low-valent metal such as yttria oxide (Y2O3) or calcium oxide (CaO) as stabilizers, followed by high-temperature sintering to form stabilized zirconia. At temperatures above 700 ℃, zirconia is an excellent conductor for oxygen ions.

01 (4)
Principle of zirconia oxygen measurement

Porous platinum (Pt) electrodes are sintered respectively on both sides of the zirconia electrolyte (ZrO2 tube). At a certain temperature, when the oxygen concentrations on the two sides of the electrolyte are different, oxygen molecules on the high-concentration side (air) are adsorbed onto the platinum electrode and combine with electrons (4e) to form oxygen ions O2−, making this electrode positively charged. O2− ions migrate through the oxygen ion vacancies in the electrolyte to the platinum electrode on the low oxygen concentration side, release electrons, and convert back into oxygen molecules, causing that electrode to become negatively charged. The reaction equations for the two electrodes are as follows:

Reference side: O₂+4e→2O²¯

Measuring side: 2O²¯ - 4e→O2₂


This generates a certain electromotive force between the two electrodes. The zirconia electrolyte, platinum electrodes, and gases with different oxygen concentrations on both sides together form the oxygen probe, namely the so-called zirconia concentration cell. The electromotive force E between the two electrodes is calculated by the Nernst equation: that is,

Solid Electrolyte Zirconia Ion Flow Detection Technology 2

In the equation:

E―Output of concentration cell, mV;

R―Ideal gas constant, 8.314 W·s/mol;

T―Absolute temperature (K);

n―Number of electron transfers (4 in this equation);

F―Faraday constant, 96,500 C;

P0―Oxygen concentration percentage of the reference gas;

P1―Oxygen concentration percentage of the gas under test.

It is the basis of zirconia oxygen measurement. When the temperature at the zirconia tube is heated to 600~1400°C, the gas on the high-concentration side uses a gas with a known oxygen concentration as the reference gas; if air is used, P0=20.6%. By combining this value with the constant terms in the formula, and considering that the actual zirconia cell exhibits thermoelectric potential, contact potential, reference potential and polarization potential, a local potential C (mV) is generated. The actual calculation formula is:

Solid Electrolyte Zirconia Ion Flow Detection Technology 3

As can be seen, if the output electromotive force E of the oxygen probe and the absolute temperature T of the measured gas can be determined, the oxygen partial pressure (concentration) P1 of the measured gas can be calculated. This is the basic oxygen measurement principle of the zirconia analyzer.


Note: The content of Detection Technology of Zirconia Analyzer is excerpted from: Mei Bo, Jin Haifeng. Principle, Maintenance and Application of Zirconia Oxygen Analyzer. Ethylene Industry (in Chinese), 2009, 21(3): 28-31.

>> Introduction to Ion Flow Sensor

Ion flow sensors are all based on the zirconia principle, and their oxygen measurement principle is referred to the Section 11.1.2. Foreign manufacturers such as Fujikura in Japan and Sensore in Austria, as well as early domestic manufacturers including Chengdu Kangda, all adopt single limiting orifices. With technological advancements and drawing on the extensive field application experience summarized by Shanghai Chang Ai, companies such as Shanghai Aici have developed porous layer oxygen sensors. This design adopts the porous ceramic substrate as the diffusion layer to control the oxygen supplied to the sensor cathode (replacing the mechanical restriction of a single hole). Due to the special material properties, uniformly distributed mesh holes are naturally formed during sintering, which are resistant to clogging.


A comparison of typical ion current sensors is shown in Table 1.

Table 1: Comparison of Common Ion Flow Sensors

Comparison Item Sensore/Fujikura AICI
Principle Ion flow 3D ion flow
Thermal Effect Glass glaze bonding technology. The glaze and zirconia substrate are different materials with different thermal expansion coefficients, making them highly susceptible to thermal stress. Cold and thermal shock easily leads to cracking at the bonded interface. Tape-casting lamination and co-firing technology, enabling uniform heating and immunity to the impact of cold and thermal shock.
Limiting current hole aperture: 10 μm Laser drilling, a form of photothermal ablation method. When a high-energy beam irradiates the surface of a material, the material rapidly heats up and vaporizes as it absorbs the light energy. Irregular spatter deposits are formed around the hole and on the inner wall, which directly affect the performance and consistency of the sensor. Porous ceramics are used; due to the special properties of the material, sintering naturally forms a uniformly distributed mesh holes.
Number of holes Single hole is prone to clogging. Naturally formed reticular porous structure, resistant to clogging.
Sensitivity T90< 60s T90< 45s
Quality guarantee 15000 hours More than 50, 000 hours
Physical object Solid Electrolyte Zirconia Ion Flow Detection Technology 4 Solid Electrolyte Zirconia Ion Flow Detection Technology 5

The current generated by OH- ion flow is proportional to the oxygen content in the sample gas. It can be seen from the above chemical reactions that if no oxygen is present, no reaction occurs and no current is generated. Therefore, the sensor theoretically has an absolute zero point. Nevertheless, similar to concentration-cell zirconia sensors, whose theoretical electromotive force in air should be zero but usually delivers a non-zero output due to materials, the signal of fuel cell oxygen sensors generally cannot reach a zero even after being supplied with high-purity nitrogen treated by deoxygenation technology, and may even produce negative signals. Since the lead at the anode is continuously converted into lead oxide, the service life of the sensor terminates once the lead electrode is completely consumed.

>> Performance Analysis

In an alkaline electrolyte solution, the reduction of oxygen to OH- at the silver cathode can be expressed by the following formula.

a01

In formula:

I -­ Current flowing through the electrodes of a galvanic cell

K­ - Constant

[O₂]  The oxygen concentration in the measured sample gas

[OH-]  The activity (effective concentration) of OH⁻ ions in the electrolyte

e - ­Base of the natural logarithm

φ- Polarization reaction potential of the silver electrode

F - Faraday constant

R - Gas constant

S - Thermodynamic temperature

This formula covers all reactions of alkaline fuel cell oxygen sensors, but can be used for the qualitative interpretation of the characteristics of fuel cell oxygen sensors.

As can be seen from the formula and Figure 6-2

① The higher the oxygen concentration, the more obvious the nonlinear relationship.

② Temperature Characteristics: The discharge current of the fuel cell oxygen sensor exhibits an exponential relationship with the thermodynamic temperature T. As the temperature rises, the discharge current increases significantly. 

Therefore, to ensure measurement accuracy, two methods can be employed: constant temperature maintenance or temperature compensation. At present, most oxygen analyzers on the market equipped with fuel cell oxygen sensors employ thermistors with a negative temperature coefficient for temperature compensation, while those using a constant-temperature method are less common.

图片 2
Figure 6-2 Relationship between oxygen concentration and output signal

③ Effect of KOH solution on fuel cell oxygen sensors

It can be concluded from the formula that OH- exhibits a negative exponential relationship with the current signal output by the sensor. Studies have shown that when the concentration of KOH solution is around 6 mol/L (mass fraction: 26.8%), the electrical conductivity reaches a maximum, meaning that the activity of OH⁻ is also at its maximum at this point. Further research indicates that when the KOH concentration is maintained within the range of 5.5~6.9 mol/L, the variation of conductivity caused by fluctuations in solution concentration and temperature is minimized. This corresponds to the smallest variation in OH⁻ activity, thereby minimizing the impact on the sensor's sensitivity. Therefore, the preparation of KOH solution for the sensor shall comply with the above principles.

④ Effect of sample gas flow rate

Variations in sample gas flow rate generally have no significant effect on the discharge current of fuel cell oxygen sensors. This is because the current signal output of the sensor is correlated with the oxygen partial pressure in the measured gas. When the sample gas flow rate changes but the oxygen content in the sample gas stays constant, the partial pressure of oxygen also remains unchanged.

>> Major Technical Specifications

Taking the CI-PC90 trace oxygen analyzer from CHANGAI Electronic Science & Technology Co., Ltd. as an example, the main technical specifications are as follows:

Sensor CI213
Accuracy 0.01~9.99ppm O₂ ±5% FS
10.0~99.9ppm O₂ ±3% FS
100~1000ppm O₂ ±2% FS
0~21.00%    O₂ ±2% FS
Repeatability 0.01~9.99ppm O₂ ±2.5% FS
10.0~99.9ppm O₂ ±1.5% FS
100~1000ppm O₂ ±1% FS
Stability 0.01~9.99ppm O₂ ±2.5% FS/7d
10.0~99.9ppm O₂ ±1.5% FS/7d
100~1000ppm O₂ ±1% FS/7d
Response Time T90<60S(25℃)
Recovery Time It takes 60 minutes to reduce the concentration from the ambient level (20.94%) to 10 ppm
Calibration Cycle One year(recommended)
Ambient Temperature 0~45℃
Ambient Humidity <80%RH
Sample Gas Pressure Normal pressure ±10% (air outlet must be vented)
Sample Gas Flow 1.5~2L/min
Sensor Service Life More than 2 years(normally use)

>> Precautions for Use

① Studies have shown that the service life of fuel cell oxygen sensors is related to the following factors: 

 Volatilization and leakage of electrolyte; 

 Passivation effect caused by lead oxide deposition from surface reaction of the lead anode metal; 

 Gas permeability and water repellency of the permeable membrane. The passivation of lead oxide is related to the measured oxygen content. The higher the oxygen concentration, the greater the anode consumption and the shorter the service life of the sensor. Therefore, it is recommended to equip a spare sensor.

② Oxygen analyzers equipped with fuel cell oxygen sensors as the detection unit require low routine maintenance. Calibration shall be performed once every six months with high-purity nitrogen (≥99.999%) and oxygen-in-nitrogen standard gas at 90% of the measurement range.

③ When the production equipment is shut down for maintenance and the analyzer is out of service, it is recommended to purge the fuel cell oxygen sensor of the analyzer with high-purity nitrogen (≥99.999%) for approximately 8 - 10 minutes, and then set the analyzer to the purge mode ( (at which point the sensor is sealed). After production equipment maintenance is completed and the analyzer is being restarted, purging the gas circuit with the measured sample gas for 3–5 minutes before switching the analyzer to the measurement mode. This operation provides two advantages: firstly, it extends the service life of the sensor; secondly, it results in faster response and stabilization times when resuming measurements. This measure is especially applicable to scenarios requiring rapid measurement, such as the production of high-purity nitrogen and high-purity argon, and CO₂ recovery in breweries.

④When storing a fuel cell oxygen sensor, place it in a nitrogen-filled protective bag and short-circuit the terminals with a shorting ring. Do not damage the protective bag during storage. The bag should only be opened when replacing the sensor. After removing the shorting ring, install the sensor into the analyzer immediately.

⑤The pressure range of fuel cell oxygen sensors is generally 35~210 kPa. If the gas supply pressure is excessively high, a pressure reducing valve must be used first to adjust the pressure within the above-mentioned safe range.

Acidic Fuel Cell Oxygen Sensor

The acidic fuel cell oxygen sensor consists of a gold cathode, a lead anode and liquid acetic acid electrolyte. It is suitable for environments where the measured atmosphere contains acidic substances (such as CO₂ and H₂S), such as the measurement of trace oxygen in CO₂ recovery at breweries and the measurement of trace oxygen under nitrogen-protected in brazing furnaces. A typical acidic fuel cell oxygen sensor is the XLT-12-333 from AII. Its schematic structure is similar to the alkaline fuel cell oxygen sensor shown in Figure 6-1, with differences only in electrode materials and electrolyte. The figure below illustrates the schematic structure of the acidic fuel cell oxygen sensor produced by CITY. Despite structural differences, both sensors share the same operating mechanism.

When oxygen in the measured gas passes through the PTFE permeable membrane (also referred to as an oxygen diffusion membrane in some literature) and enters the fuel cell, the following redox reactions occur at the electrodes.

The main difference between alkaline and acidic fuel cell oxygen sensors lies in their electrolytes.This design is intended to accommodate various application scenarios.With the advancement of technology, some companies have developed fuel cell oxygen sensors using neutral electrolytes, such as the CI213 model from Changai, which is suitable for applications where the measured atmosphere contains acidic or alkaline gases.

图片 3
Figure 6-3 Schematic Structure of CITY Acidic Fuel Cell Oxygen Sensor
Cathodic reduction reaction O₂+2H₂O+4e-→4OH-
Anodic oxidation reaction 2Pb+ 4OH-→2 PbO+2H₂O+4e-
Overall cell reaction O₂+ 2Pb→2 PbO

Electrolytic Cell Oxygen Analyzer

Essentially, an electrolytic cell converts electrical energy into chemical energy. The electrolytic cell oxygen sensor belongs to the electrolytic cell category. Therefore, in principle, its electrochemical reaction requires an external power supply for normal operation. Compared with fuel cell oxygen sensors, its anode is non-consumable and generally does not need replacement. Electrolytic cell oxygen sensors are mainly used for trace oxygen measurement, with a detection limit down to the ppb level (currently, the vast majority of fuel cell-type oxygen sensors used for trace oxygen measurement can only achieve the ppm level). A typical electrolytic oxygen analyzer is the Delta F trace oxygen analyzer manufactured by GE (see Figure 6-4 for the schematic structural diagram of the sensor). Its sensor is based on the coulometric electrolysis principle. A DC voltage of approximately 1.3 V is applied to the electrolytic cell to supply energy for redox reactions. When trace oxygen in the sample gas passes through the permeable membrane into the cathode, oxygen molecules are reduced to OH⁻ at the cathode. With the aid of KOH electrolyte, OH⁻ migrates to the anode where an oxidation reaction takes place to generate oxygen, which is then discharged.

Cathodic reduction reaction O₂+2H₂O+4e-→4OH
Anodic oxidation reaction 4OH-→O₂+2H₂O+4e

As can be seen from the above electrode reaction equations, there is no consumption of the electrolytic cell or electrodes. Therefore, users do not need to replace the electrodes or electrolytic cell during operation; they only need to periodically replenish the distilled water and electrolyte (the electrolyte decreases due to natural evaporation). This is different from the aforementioned fuel cell oxygen sensors, which generally need to be replaced every 1 to 2 years.

When introducing alkaline fuel cell-type oxygen sensors, it is emphasized that they must not be used in applications where the measured gas contains acidic components. The Delta F electrolytic oxygen sensor uses an alkaline KOH solution as its electrolyte. To overcome interference caused by acidic gases and prevent electrode corrosion, a pair of Stab-EL auxiliary electrodes is designed inside the sensor. The function of these auxiliary electrodes is to remove these harmful gases after the sample gas containing acidic gases enters the electrolytic cell, thereby preventing damage to the sensor and ensuring the accuracy of the analyzer's readings.

图片 4

Figure 6-4 Schematic diagram of the Delta F trace oxygen sensor

예전
수소 에너지 응용 분야
추천 콘텐츠
저희에게 연락하세요
창아이(CHANG AI)는 가스 분석 및 감지 분야의 세계적인 선도 첨단 기술 기업으로, 고객에게 세계 최고 수준의 가스 분석 및 감지 제품과 원스톱 솔루션을 제공하기 위해 노력하고 있습니다.
연락처 정보
팩스: +86-21-33275656
전화: +86-21-51692285 / +86-21 400 700 8817
추가: 중국 상하이 민항구 신롱루 1333항 Qibao WanKe 국제센터 97호. 201101
Customer service
detect